YSI MultiLab 4010-3 User Manual
Page 93
MultiLab 4010-3
Settings
ba76143e03
08/2013
93
Calibration / Calibra-
tion data storage /
Display
-
Displays the last calibration records
(max. 10)
Calibration / Calibra-
tion data storage /
Output to USB flash
drive or printer
-
Outputs the stored calibration data to
the USB-A interface (USB memory
device/USB printer)
Calibration /
Calibration data stor-
age / Output to
RS232/USB
-
Outputs the stored calibration data to
the USB-B interface (PC)
Calibration /Calibra-
tion interval
1 ... 150 ...
999 d
Calibration interval for the IDS con-
ductivity sensor (in days).
The meter reminds you to calibrate
regularly by the flashing sensor sym-
bol in the measuring screen.
Type
Used measuring cell
Cal
Measuring cells whose cell constant
is determined by calibration in the KCl
control standard solution.
Calibration range:
0.450 to 0.500 cm
-1
The currently valid cell constant is dis-
played in the status line.
man
Freely adjustable cell constant in the
range 0.450 ... 0.500 cm
-1
.
Man. cell const.
0.450 ...
0.475 ...
0.500 cm
-1
Display and setting options for the cell
constant.
Temp. comp. (TC) /
Method
nLF
lin
off
Procedure for temperature compen-
sation (see section 9.2 T
This setting is only available for the
measured parameters,
ϰ and ρ.
Temp. comp. (TC) /
Linear coeff.
0.000 ...
2.000 ...
3.000 %/K
Coefficient of the linear temperature
compensation.
This menu item is only available when
the linear temperature compensation
is set.
Temp. comp. (TC) /
Reference temp.
20 °C
25 °C
Reference temperature
This setting is only available for the
measured parameters,
ϰ and ρ.
Multiplier for TDS
0.40 ... 1.00
Factor for TDS value
Menu item
Possible
setting
Explanation