MultiLab 4010-3

pH value

ba76143e03

08/2013

33

4.

Thoroughly rinse the sensor with deionized water.

5.

Immerse the pH sensor in buffer solution 1.

6.

For measurements without temperature sensor
(e.g. when using an IDS adapter

)

:

Enter the temperature of the buffer with <><>.

7.

Start the measurement with .
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter flashes.

8.

Wait for the measurement with stability control to be completed or ter-
minate the stability control and take over the calibration value with
.
The pH value of the buffer solution is displayed.