beautypg.com

4 calibration standards, 5 calibration data, Calibration standards – YSI MultiLab 4010-3 User Manual

Page 53: Calibration data

background image

MultiLab 4010-3

Ion concentration

ba76143e03

08/2013

53

7.2.4 Calibration standards

Use two to seven different standard solutions. The standard solutions have to
be selected in either increasing or decreasing order.

7.2.5 Calibration data

Displaying the cali-

bration data

The calibration record of the last calibration is to be found under the menu item,
Calibration / Calibration record. To open it in the measured value display, press
the key.

The calibration records of the last 10 calibrations are to be found in the menu,
Calibration / Calibration data storage / Display. To open the Calibration menu,
press the key in the measured value display.

Select the unit of the standard solution and measurement result in
the ISE setup/Unit menu.

Standard solution (Std 1 - 7) Values

Unit [mg/l]

0.010 ... 500,000

Unit [mol/l]

0.100 ... 5,000 µmol/l
10.00 ... 5,000 mmol/l

Unit [mg/kg]

0.010 ... 500,000

Unit [ppm]

0.010 ... 500,000

Unit [%]

0.001 ... 50,000

The measurement precision is also dependent on the selected
standard solutions. Therefore, the selected standard solutions
should cover the value range expected of the subsequent concen-
tration measurement.

If the measured electrode potential is outside the calibrated range,
the [ISEErr] warning is displayed.

The calibration record is automatically transmitted to the interface
after calibrating.