MultiLab 4010-3

Settings

ba76143e03

08/2013

91

ISE setup /
Ion type

Ag, Br, Ca, Cd, Cl,
CN, Cu, F, I, K, Na,
NO3, Pb, S, NH3,
NH4*, CO2, ION

Selection of the ion type to
be measured.

An ion that is not included in
the list can be measured
with the setting, ION.

* Measuring with the NH 500 electrode:

The NH4 setting is not suitable for the gas-sensitive
electrode NH 500.
Select the following settings:
Ion type "ION", Valency "-1".

ISE setup /
Unit

mg/l
µmol/l
mg/kg
ppm
%

Selection, with which unit
the measurement result and
calibration standards should
be displayed.

ISE setup/
Valency

-8 ... +8

Set the valence (Valency)
and molar weight (Molar
mass
) for the ion (only with
Unit: mg/kg, ppm, %)

ISE setup/
Molar mass

1 ... 300 g/mol

ISE setup/
Density

0.001 ... 9.999 g/ml or
kg/l

Adjustable density of the
test sample (only with Unit:
mg/kg, ppm, %)

Method

Standard addition
Standard subtraction
Sample addition
Sample subtraction
Blank value addition

Selection of the available
measuring methods.

Start method

Start measurement with the
selected method.

Menu item

Possible setting

Description