DT

Delta Time

[ 2500sub0B ]

Set the Delta Time in milliseconds [ms].

During MT and TT timeframes "sub-averages" will be

calculated by the system over the time DT. If a sub-average is outside the re-trigger window,

the re-trigger function is automatic started.

Master (PC / PLC) sends

Device responds

Result

DT T+00000

Request: DT=00000

DT_50 OK

Setting: TT=50ms

Factory default: 00000

TW

Tare Window

[ 2500sub06 ]

Set the Tare Window in in counts (digits) without decimal point. Tare window (TW) allows an

automatic Tare update. If TW = 0 this function is not active. If TW = 100, this means a new

tare value will be taken when the net average weight of an empty scale is within 100 counts

or division of zero. The new average tare value is calculated over the average tare time

defined by TI. If the tare average is outside tare window, the tare will not be updated.

Master (PC / PLC) sends

Device responds

Result

TW T+00000

Request: TW=00000

TW_100 OK

Setting: TW=100d

Factory default: 00000

TI

Tare Time

[ 2500sub07 ]

Set the Tare Time in milliseconds [ms].

During the Tare Time a “tare-average” will be

calculated by the system.

Master (PC / PLC) sends

Device responds

Result

TI T+00000

Request: TI=00000

TI_200 OK

Setting: TI=200ms

Factory default: 00000


HT

Hold Time

[ 2500sub05 ]

Set the Hold Time in milliseconds [ms].

During the Hold Time the weight value must be over

setpoint that a digital output can be switched. This means that a short signal peak will not

lead to the switching of a digital output.

Master (PC / PLC) sends

Device responds

Result

HT T+00000

Request: HT=00000

HT_200 OK

Setting: HT=200ms

Factory default: 00000