Page Manual MSC-64

32

AZ

Absolute zero point calibration

[ 2300sub02 ]

The command AZ is used as reference point for all weight calculations and will setup in

mV/V. Permitted values are ± 32000 (= ± 3.2000 mV/V).

Master (PC / PLC) sends

Device responds

Result

AZ

Z+0.0005

Request :

Zero point @ 0.0005mV/V

CE

E+00017

(example)

Request: TAC-counter CE 17

CE_17 OK

Calibration sequence active

AZ_00500 OK

Setting:

new: Zero point @ 0.0500 mV/V

Factory default: 00000 @ 0.0000 mV/V input signal.


AG

Absolute gain calibration

[ 2300sub01 ]

The command AG is used as absolute gain (or measuring range) for all weight

calculations and will setup in mV/V. Permitted values are ± 32000 (= ± 3.2000 mV/V).

Master (PC / PLC) sends

Device responds

Result

AG

G+2.0000

Request :

Meas. range 2.000 mV/V

CE

E+00017

(example)

Request: TAC-counter CE 17

CE_17 OK

Calibration sequence active

AG_19998_90900

OK

new: Measuring range

1.9998 mV/V @ 90900 d

Factory default: 200000 @ 2.0000 mV/V input signal.

If the AG value has to be set at 2.0000 mV/V, but the load cell capacity is quoted at

2.2 mV/V, you will have to ratio the load cell capacity down.

Example : A 100kg load cell with an Output at Rated Load of 2.2 mV/V would be equivalent

to 90.9 kg @ 2 mV/V (exact 1.9998 mV/V)


ZT

Zero

tracking

[ 2100sub12 ]

This command enables or disables the zero tracking. Parameter = 0 disables the zero

tracking and parameter = 1 enables the zero tracking. Issuing the command without any

parameter returns the current ZT value.

Master (PC / PLC) sends

Device responds

Result

ZT Z:001

Request : ZT status (ON)

CE

E+00017

(example)

Request: TAC-counter CE 17

CE_17 OK

Calibration sequence active

ZT_0 OK

Setting: ZT = OFF