Flintec MCS-64 CANbus User Manual

Page 19

background image


Manual MCS-64 Page 19

Index

Sub-
index

Name

Type

Attri-
bute

Default-value

Meaning

2300 -
233F

for
LDM
0..63












0
1
2
3
4
5
6
7
8
9
10
11
12

Number of entries.
Absolute gain
Absolute zero
Calibrate enable
Calibrate gain
Set calibration point B
Set calibration point A
Calibrate max
Calibrate min
Calibrate save
Calibrate zero
Decimal point
Display step size

UI8
I32
I32
I32
I32
I32
I32
I32
I32
I32
I32
I32
I32

ro
rw
rw
rw
rw
rw
rw
rw
rw
rw
rw
rw
rw

12

Number of calibration parameters.
Absolute gain calibrate (TAC protected)
Absolute zero calibrate (TAC protected)
Calibrate enable (enables TAC when the TAC is written)
Calibrate gain (TAC protected)
Set calibration point B
Set calibration point A
Calibrate max (TAC protected)
Calibrate min (TAC protected)
Calibrate save (TAC protected)
Calibrate zero (TAC protected)
Decimal point (TAC protected)
Display step size (TAC protect)

23FF














0

Number of entries.
As for 2300-233F

UI8

ro
wo

12

Number of calibration parameters.
As for 2300 – 233F, except this is WRITE ONLY and the settings are
broadcasted to all LDMs.