Flintec MCS-64 CANbus User Manual

Page 33

background image

Manual MCS-64 Page 33

Zero tracking will be performed only on results less than +/-0.5 d at a rate of 0.4 d/sec, where

d = display step size (see DS command). The zero can only be tracked to +/- 2% of

maximum (see CM command).

In non legal for trade applications you can exceed the limit +/- 2%. If you require for example

a zero tracking range of +/- 10 d then you have to set ZT to be 100 where d is the division

size (which means d * 10).

Factory default: ZT=0


FD

Factory default settings

[ 2006sub02 ]

This command puts the LDM88.x back to a known state. The data will be written to the

EEPROM and the TAC will be incremented by 1.


Note:

All calibration and setup information will be lost by issuing this command!

Master (PC / PLC) sends

Device responds

Result

CE

E+00017

(example)

Request: TAC-counter CE 17

CE_17 OK

Calibration sequence active

FD OK

Factory default setting


CS

Save the calibration values

[ 2004sub02 ]

This command results in the calibration values being saved to EEPROM, and causes the

TAC to be incremented by 1.

Master (PC / PLC) sends

Device responds

Result

CE

E+00017

(example)

Request: TAC-counter CE 17

CE_17 OK

Calibration sequence active

CS OK

Calibration values saved

The CS command saves all of the calibration group values, as set by CZ, CG, CM, DS, DP

and ZT. The command returns ERR and has no updating action unless it is preceded by the

CE_XXXXX.