Specifications, N -b u s – Flintec MCS-08 User Manual
Page 7
MCS-08 Technical Manual, Rev. 1.04 December 2012
Page 7 of 52
Figure 3.2
– MCS-08 bus system
A basic configuration of type MCS-08 Mult Channel System is shown in Figure 3.2. In this configuration the
gateway module is the master of the internal N-bus and it simultaneously acts as a slave for the external bus.
The gateway connects the external network with the locally installed MCS-08 hardware modules and transmits
commands and responses to and from the external bus. It scans all hardware modules for their status and then
transmits this status information continuously to the external bus controller.
3.2. Specifications
1-channel A/D Converter: Type MCS-08AD
A/D Converter
Type
24-bit Delta-Sigma ratiometric with integral analogue and digital filter
Analogue input range
0 mV to 18 mV (unipolar) or -18 mV to +18 mV (bipolar), switchable
Linearity
< 0.0015 % FS
Temperature coefficient
< 2 ppm/°C
Min. input sensitivity
0.1 µV/d
Conversion rate
Up to 800 measurement values per second
Internal resolution
Up to 8 million counts
External resolution
Up to 100000 counts (weight value, force, torque) respective
1 million raw counts (unipolar) respective 2 million raw counts (bipolar)
Calibration and Weighing Functions:
Calibration
Electronic calibration without test weights (eCal) or calibration by test weights
Digital filter
10 step adjustable digital adaptive filter
Weighing functions
Tare, zero, auto zero tracking, motion detection, auto-zero at power-up, save
tare at power-off, increased resolution
Load cells:
Excitation:
5 V DC at
58...1200 Ω, max. 100 mA
Number of load cells:
Up to 6 load cells à 350 Ω or 18 load cells à 1100 Ω in parallel
Connection:
4- or 6-wire technique, cable length 250 m/mm² for 6-wire connection
Communication and Setup:
Setup & calibration
By PC software via gateway module, backup data stored on PC
Response time
< 4 ms (delay after each read or write command)
Power supply:
DC power supply
10 to 28 VDC, < 200 mA, not galvanically isolated
Environment and Enclosure:
Operating temperature
Between -10 °C and +40 °C at maximum 85% RH, non-condensing
Enclosure & protection class Polyamide, for DIN-rail mounting, protection class IP20
Dimensions & weight
99 x 22.5 x 114.5 mm (L x W x H), weighs approx. 100 g
N -b u s
N -b u s
E xte rn a l b u s
N -b u s
N -b u s
E xte rn a l b u s