beautypg.com

Flintec MCS-08 User Manual

Page 24

background image

MCS-08 Technical Manual, Rev. 1.04 December 2012

Page 24 of 52

Zeroing MCS-08AD [1] at register 41063.
Request:

01, 10, 04, 26, 00, 01, 02, 00, 01, 25, 56

Answer:

01, 10, 04, 26, 00, 01, E1, 32

MCS-08AD [1] is zeroed.

Taring MCS-08AD [0] at register 41062.
Request:

01, 10, 04, 25, 00, 01, 02, 00, 02, 65, 64

Answer:

01, 10, 04, 25, 00, 01, 11, 32

MCS-08AD [0] is tared.

Taring MCS-08AD [1] at register 41063.
Request:

01, 10, 04, 26, 00, 01, 02, 00, 02, 65, 57

Answer:

01, 10, 04, 26, 00, 01, E1, 32

MCS-08AD [1] is tared.

Zero Calibration of MX08 AD [0].
Request:

01, 03, 04, 66, 00, 01, 65, 25 (Read status; Ready status is mandatory)

Answer:

01, 03, 02, 00, 01, 79, 84 (MCS-08AD [0] is in ready status; zero calibration

can be performed)
Request:

01, 10, 04, 63, 00, 01, 02, 00, BC, EB, B2 (Zero calibration command)

Answer:

01, 10, 00, 6D, 00, 01, 90, 14

Request:

01, 03, 04, 66, 00, 01, 65, 25 (Read status and wait for ready status)

Answer:

01, 03, 02, 00, 01, 79, 84 (If “zeroing” changed to “ready” status, zero

calibration was performed successfully.)

Span Calibration of MCS-08AD [0].
Request:

01, 03, 04, 66, 00, 01, 65, 25 (Read status; Ready status is mandatory)

Answer:

01, 03, 02, 00, 01, 79, 84 (MCS-08AD [0] is in ready status; span calibration

can be performed)
Request:

01, 10, 04, 63, 00, 03, 06, 00, DC, 00, 00, 0B, B8, D6, 90 (Span calibration

command with 3000 (0x0BB8 hex) span value)
Answer:

01, 10, 00, 6D, 00, 03, 11, D5

Request:

01, 03, 04, 66, 00, 01, 65, 25 (Read status and wait for ready status)

Answer:

01, 03

, 02, 00, 01, 79, 84 (If “span calibr.” changed to ready status, span

calibration was performed successfully.)

The table below is used to access MCS-08AD modules.
The starting register address 40001 is valid for N-bus address [0].

Address R/W

Word Command

Definition

40001*

R

2

Weight / Force / Count Data

40003*

R

1

Module Status of
MCS-08AD

D0

0

– System Ready

1

– System Busy

D1

0

– Error

1

– Data ok

D2

0

– Weight Stable

1

– Weight unstable

D3

0

– Gross Mode

1

– Net mode

D4

0

Not used

D5

0

– Weight / Force

1

– Count Mode

D6…D11

Not used

D12

0

– Out of zero range 1 – Weight is in zero range

D13
D14
D15

Error Code

0

No Errors

1

ADC out of range

2

ADC over range

3

ADC under range

4

System error

5

In setup mode

6

Low/High voltage det.

40004*

R

2

Tare weight

40006*

R

2

Gross weight

40008*

R

1

Status

Motion, Net mode, Data ok, (image of register 40003)

40009*

R/W

1

Control

0

None

1

Zero

2

Tare

3

Clear