MCS-08AD modules are state-of-the-art strain gauge load cell signal digitizers. These modules are used for
any type of process weighing and force measurement including tank and silo weighing, dynamic weighing,
check weighing, filling, tension /compression force measurement etc.
6.1. Front View
There are 3 status LEDs on the front panel which indicate the operational module status (Refer to table 6.1).
The setup switch on front panel of the module is used for N-bus addressing without PC (Refer to chapter 5.6)
and for diagnostics (Refer to chapter 15).
When the error LED is ON, the other two LED indicate the error type (Refer to chapter 14 for details). Load cell
connection and power supply terminals are located at the front of the DIN rail mount module (See figure 6.1).
Status LEDs
Load Cell Connector
Setup Switch
Power Supply Terminal
Figure 6.1
– Front view
The meanings of these LEDs in operation are described below.
LED
Operational Mode
Symbol
Name
Weight / Force
Count
Stable
Stable
Unstable
Off for 0.3 seconds in 2 seconds
period (No stable indication)
Centre
of Zero
in the centre of zero range
( -0.25 e < w < 0.25 e )
Out of centre of zero range
Always off
( No centre of zero indication )
Err
Error
( * )
ADC conversion error
Digital processing error
No error
ADC conversion error
Digital processing error
No error
Off
On
Flashing Off for 0.3 seconds
( * ) : Refer to the error table in chapter 14
Table 6.1