Back up settings and calibration data – Flintec MCS-08 User Manual
Page 22
MCS-08 Technical Manual, Rev. 1.04 December 2012
Page 22 of 52
4. Now change the address number on the Gateway Module by pressing the gateway switch (
YELLOW
) until it
reached an address you want to use for an MCS08AD or MCS08IO module.
5. To assign the selected address number to an A/D Converter or Digital I/O module press the setup switch on
the selected module (
BLUE
).
Please note: The MCS08 allows up to 8 A/D and 8 I/O in one system. A/Ds and I/Os
use the same address range (0 to 7). In consequence the same address can be
assigned to one A/D Converter AND one Digital I/O Module.
6. After assigning all modules to a defined address, press and hold the gateways switch for 5 seconds. The
system will close the addressing mode and change to normal mode
Note:
Do not de-energise MCS-08 while the system is in the addressing mode. If a power failure has
happened, restart the system and repeat the whole addressing.
LED and Setup Switch Descriptions in the N-Bus Addressing Mode:
Gateway module
Description of LED
Description of Setup Switch
Pwr
Lnk
Err
Address number is [0]
Increase to address number [1]
Address number is [1]
Increase to address number [2]
Address number is [2]
Increase to address number [3]
Address number is [3]
Increase to address number [4]
Address number is [4]
Increase to address number [4]
Address number is [5]
Increase to address number [6]
Address number is [6]
Increase to address number [7]
Address number is [7]
Jump to address number [0]
Address number is higher than [7]
MCS-08AD or MCS-08IO
Description of LED
Description of Setup Switch
Err
Defined N-bus address number and
gateway‟s selected address number
are identical
Assigns the gateway‟s selected address
number to its N-bus address number.
Defined N-bus address number and
gateway‟s selected address number
are different
Increase to address number [2]
Off
On
Flash
Table 5.2
– N-bus addressing via setup switch
5.9. Back up Settings and Calibration Data
The parameter settings and the calibration coefficients can be saved to a backup file after reading these data
from the MCS-08 modules. This backup file can be re-written to MCS-08 modules after opening it by xFace.
The backup feature gives a service advantage to MCS-08 systems. We propose to burn these files to CD /
DVD and save this backup CD / DVD in the control cabinet additionally to a backup file on your PC.
Important note: The previous parameter settings and calibration coefficients in a MCS-08 module will get lost
after loading a backup to this module.