MCS-08 Technical Manual, Rev. 1.04 December 2012

Page 21 of 52

5.7. Digital I/O Test

Figure 5.10

– MCS-08IO

Status

Input / Output Test: This block allows the user to check I/O status. The I/O test button is used to
test outputs as seen in this figure.

The active MCS-08IO modules are shown with their own N-Bus address block. You can follow the input and the
output status and you can switch the outputs. The I/O test property is very useful for checking the output states.
Warning: Please be careful to avoid damaging your process during the output test.

5.8. N-Bus Addressing via Setup Switch

MCS-08 systems which are energized for the first time require the N-bus addressing for each A/D Converter
module and each Digital I/O module. One way is the addressing by the xFace which is described above.
Another way is the addressing (up to address 7) via the setup switch.

Figure 5.11

– Example


Step by step manual procedure:
1. Unplug all communication cables from the Gateway Module (RS232 and BUS plug; See

GREEN

).

2. First energize the modules installed on the N-Bus.
3. Press the setup switch of the gateway module for 5 seconds. (The switch is behind the hole in the front; see

YELLOW

). The three LEDs of the gateway module (

RED

) will flash 4 times and then indicate an N-Bus

address number. You can change the address number by pressing the setup switch (

YELLOW

) slowly.

While changing the address in the Gateway module, the LEDs in the other modules (

PINK

) will change

sometimes.