Flintec MCS-08 User Manual
Page 39
MCS-08 Technical Manual, Rev. 1.04 December 2012
Page 39 of 52
Figure 9.3
– MCS-08MB gateway setup
You will see the Modbus setup in this tab.
Data Format
Modbus RTU
Baud Rate
One of the following baudrate will be selected for the RS-485 communication port.
1200
2400
4800
9600 (default setting)
19200
38400
57600
Data Length and Parity
The data length and parity can be selected as 8 None 1 (default), 7 Odd 1 or 7 Even 1.
RS485 Address
Data Format Modbus RTU: The Modbus slave address range is 01 (default) to 31.
If you enter 0, the instrument will operate without address data.
Displayed Gateway Information
External Bus: Modbus RTU
Firmware Version: Revision number of the firmware
Firmware Date: Release date of the firmware
Hardware Version: Revision number of the main printed circuit board
Serial Number: The module‟s serial number
Gateway Status: Indicates the system OK status
Modbus RTU Data Structure
Please refer to chapter 5.11