Address R/W
Word Command
Definition
40010*
R/W
1
Calibration
0
None
188
Adjust Zero Calibration
220
Adjust Span Calibration (First load 40011 with span test
weight value)
40011*
R/W
2
Span Calibration Value
40013*
R
1
Calibration
Status
D0 .. D7
Calibration
Process
Status
1
Ready for calibration
3
Zero calibration in process
4
Span calibration in process
9
Error (Refer to D8 ... D15 )
D8 .. D15
Calibration
Errors
1
Calibration Timeout
- Restart calibration
2
ADC Error
- Re-energize the instrument
- If seen again, change the board.
3
Instrument can not be calibrated
- Check load cell cable
- Re-energize the instrument
34
Instrument can not be calibrating
- Load cell signal is very low or too high
35
Calibration Error
- Calibration test weight is too small
- Increase calibration weight value (40011)
- Check load cell connections
37
Scale unstable
- Wait until scale becomes stable
- Check grounding wiring
40014*
R/W
1
Operation Mode
Selector
0
Count Mode Unipolar
1
Count Mode Bipolar
2
Force Mode Unipolar
3
Force Mode Bipolar
4
Weight Mode Unipolar
40015*
R/W
1
mV operation
in Count Mode
0
5 mV
1
10 mV
2
15 mV
3
18 mV
40016*
R/W
1
Digital filters
0
Fast
1
2
3
4
5
Medium
6
7
Default
8
9
Slow
N-bus address
0
1
2
3
4
5
6
7
*Starting register address
40001
40101
40201
40301
40401
40501
40601
40701
Address R/W
Word Command
Description
41001
R
1
System Status of
MCS-08AD and
MCS-08DP
D0
00: No module found at address [0]
D1
01: Module at address [0] is active
D2
00: No module found at address [1]
D3
01: Module at address [1] is active
D4
00: No module found at address [2]
D5
01: Module at address [2] is active
D6
00: No module found at address [3]
D7
01: Module at address [3] is active
D8
00: No module found at address [4]
D9
01: Module at address [4] is active