Page 49 of 52

13. E

RROR

T

ABLE

The MCS-08 modules have been designed as very reliable and virtually error free instruments. However if an
error occurs do not attempt to repair the equipment before you understand what caused the error. Note the
problems you have with your instrument and the error messages shown by the LEDs located on the front panel.
Then try to solve the problem according to the error tables given below.

Error Code

Description

Actions to undertake / Possible cause

Err

ADC error

− Re-energize the module
− Instrument could be defective

Overload

− Check the load on the affected MCS-08AD module
− Check the calibration of the affected MCS-08AD module
− Load cell or module may be defective
− Re-energize the module. If seen again, replace the affected module

Underrange

ADC out

− Check the load on the affected MCS-08AD module
− Check the calibration of the affected MCS-08AD module
− Load cell or module may be defective
− Re-energize the module. If seen again, replace the affected module

System error

− Re-energize the module. If seen again, replace the affected module

Configuration error

− Re-address the module on the N-Bus (Refer to chapter 5.3 or 5.8)

N-bus
communication error

− Check the gateway module if it is in the system and energized
− Check the N-bus connectors if they are properly installed on the rail

High voltage detected

− Check the power supply if the voltage is within the required range

Low voltage detected

Off

On

Flashing

Table 14.1

– Error table for MCS-08AD module

Error Code

Description

Actions to undertake / Possible cause

Pwr Lnk Err

System error

− Re-energize the instrument.
− Module may be defective.

Configuration error

− Re-address the module on the N-Bus (Refer to chapter 5.3 or 5.8)

N-bus
communication error

− Check the gateway module if it is in the system and energized
− Check the N-bus connectors if they are properly installed on the rail

Off

On

Flashing

Table 14.2

– Error table for MCS-08IO and MCS-08 gateway modules

Message

Description

Actions to undertake / Possible cause

- - - - -

N-bus
communication error

− Re-address the instruments on the N-Bus (Refer to chapter 5.3 or 5.8)
− Check the gateway module if it is in the system and energized
− Check the N-bus connectors if they are properly installed on the DIN-rail

PROG

Programming

− MCS-08 is in setup mode

SYS

System error

− Re-energize the module. If seen again, replace the affected module

ADC

ADC error

− Check the load on the affected MCS-08AD module
− Check the calibration of the affected MCS-08AD module
− Load cell or module may be defective
− Re-energize the module. If seen again, replace the affected module

OVER

Overload

− Check the load on the affected MCS-08AD module
− Check the calibration of the affected MCS-08AD module
− Load cell or module may be defective

UNDER

Underrange

Table 14.3