B.3 set up flowcharts 7, B.3.1 zero set up, menu 1.1 - 1.4 7 – Flintec DAS 72.1 User Manual

Page 7

background image

B.3

SET UP FLOWCHARTS

B.3.1 Zero set up, Menu 1.1 - 1.4

Manual DAS 72.1

Page 7

Store Tare value, setting ON or OFF
OFF: Tare value deleted at power off
ON: Tare value stored non-volatile (max. 1 000 000 times)

0

T

Inp.1

0

T

Inp.1

CALIBRATE OR ADJUST THE ZERO POINT

Remark:

(CONVENTIONAL WEIGHING SYSTEM)

Display shows the actual input signal in mV/V.
Press the Enter key to store the zero.

scale should be unloaded.

CALIBRATE THE ZERO POINT FROM THE LOAD CELL mV/V READING

Use the UP/DOWN & MOVE RIGHT keys to set the mV/V reading at which the
unit should read zero.

SET UP

0

T

Inp.1

T

T

T

T

T

T

Inp.2

Inp.2

0

T

Inp.1

0

0

T

T

Inp.1

Inp.1

Press the UP or DOWN key
for more than 3 seconds to enter the Set-up Menu

0

T

Inp.1

T

T

T

Inp.2

Note:
Recessed Enable Switch
(Enables changes to be made
to important parameters.
Must be pressed before
attempting to change parameters
1.1 to 1.3, 2.1 to 2.3 and 3.1 to 3.3)

Automatic Zero Tracking
Settings: 0 ... 255 d
Function is swiched off at setting 00 000.
Zero Setting with zero key or by external signal is not possible (Menu 6).

Function is swiched on at setting 00 001 or higher, (max 00 255).
Zero Setting with zero key or by external signal (Menu 6).
Setting 00 001 sets a zero range of +/- 2 % of Upper Display Limit (Menu 3.1),
setting 00 002 up to 00 255 sets a range of 2 d up to 255 d (independant of
decimal point setting).

0

T

Inp.1

T

T

T

T

T

T

Inp.2

Inp.2

0

T

Inp.1

T

T

T

Inp.2

0

T

Inp.1

EX-Modus, setting ON or OFF
OFF:Load cell connection without Zener barrieres
ON: Load cell connection with Zener barrieres,

extended error band for ERR 5