Absolute zero point calibration

The command AZ is used as reference point for all weight calculations and will setup in mV/V. Permitted values
are ± 32 000 (= ± 3.2000 mV/V).

Master ( PC / PLC ) sends

DAS 72.1 responds

Result

AZ

¿

Z+0.0005

Request: Zero point @ 0.0005 mV/V

CE

¿

E+00017 (example)

Request: TAC-counter CE17

CE_17

¿

OK

Calibration sequence active

AZ_00500

¿

OK

new: Zero point @ 0.0500 mV/V

Factory default: 00 000 @ 0.0000 mV/V input signal.

AG

Absolute gain calibration

The command AG is used as absolute gain (or measuring range) for all weight calculations and will setup in
mV/V. Permitted values are ± 32 000 (= ± 3.2000 mV/V).

Master ( PC / PLC ) sends

DAS 72.1 responds

Result

AG

¿

G+2.0000

Request: Meas. range 2.000 mV/V

CE

¿

E+00017 (example)

Request: TAC-counter CE17

CE_17

¿

OK

Calibration sequence active

AG_17500_10000

¿

OK

new: Measuring range

1.750 mV/V @ 10 000 d

Factory default: 10 000 @ 2.0000 mV/V input signal.

ZT

Zero tracking

This command enables or disables the zero tracking. Parameter = 0 disables the zero tracking and
parameter = 1 enables the zero tracking. Issuing the command without any parameter returns the current
ZT value.

Master ( PC / PLC ) sends

DAS 72.1 responds

Result

ZT

¿

Z: 001

Request: ZT-Status

CE

¿

E+00017 (example)

Request: TAC-counter CE17

CE_17

¿

OK

Calibration sequence active

ZT 0

¿

OK

Setting: ZT = OFF

Zero tracking will be performed only on results less than +/-0.5 d at a rate of 0.4 d/sec, where d = display
step size (see DS command).
The zero can only be tracked to +/- 2% of maximum (see CM command).

Factory default: ZT=0