Flintec DAS 72.1 User Manual
Page 49

A2
Request / Set the Base for logic output 1,2 or 3
The A1 command defines the Base on which the setpoint value acts. If A1 is set to “0” then setpoint 1 acts
on the unfiltered gross weight. If A1 is set to “1” then setpoint 1 acts on the unfiltered net weight.
A1 = 0
Not filtered gross weight
A1 = 1
Not filtered net weight
A1 = 2
Peak value
A1 = 3
Average value
A1 = 4
Hold value
A1 = 5
Peak to Peak value
A1 = 6
Valley value
A1 = 7
Error message for ERROR 4/5
A1 = 8
Output diabled
Request / Set Base for logic output 1
Master ( PC / PLC ) sends
DAS 72.1 responds
Result
A1
¿
0+00000
Allocation gross weight
A1_1
¿
OK
Allocation net weight
Similarly to read or set the logic output 2/3 Base, use A2/A3 instead of A1.
NOTE: All changes to the setpoint settings have to be stored in EEPROM using the SS command. See
section 4.13.
Manual DAS 72.1
Page 49