beautypg.com

Flintec DAS 72.1 User Manual

Page 39

background image

Mode 1 Characteristic (FIR-Filter)

FL

Settling time to

0.1%

(ms)

3 dB Cut-off

(Hz)

20 dB damping

at frequency

(Hz)

40 dB damping

at frequency

(Hz)

Damping in the

stopband

(dB)

Stopband

(Hz)

Output rate

max.

(samples/s)

0

no filtering

**

600

1

47

19.7

48

64

>90

>80

600

2

93

9.8

24

32

>90

>40

300

3

140

6.5

16

21

>90

>26

200

4

187

4.9

12

16

>90

>20

150

5

233

3.9

10

13

>90

>16

120

6

280

3.2

8

11

>90

>13

100

7

327

2.8

7

9

>90

>11

85.7

8

373

2.5

6

8

>90

>10

75

** Antialiasing filter 17 Hz @ 60 dB/decade

Attention:
In mode 1 the output rate is dependant on the selected filter level (FL).

UR

Set the update rate (average building)

This command will define over how many measurements, from the preceeding IIR or FIR filter, an average
will be calculated. The average will be calculated over 2

UR

samples. Permitted values 0...7 (see table below)

UR

0

1

2

3

4

5

6

7

number of samples

1

2

4

8

16

32

64

128

Check / adjustment update rate:

Master ( PC / PLC ) sends

LDU XX.X responds

Result

UR

¿

U+0001

Request: average over 2 samples

UR_4

¿

OK

Setting: average over 16 samples

Factory default: 0 [no average, = 600 samples/s]

Remark to Mode 1

Dependency Output Rate - averaging UR - Filter FL

UR

Output Rate samples/s

FL 0

FL 1

19.7 Hz

FL 2

9.8 Hz

FL 3

6.5 Hz

FL 4

4.9 Hz

FL 5

3.9 Hz

FL 6

3.2 Hz

FL 7

2.8 Hz

FL 8

2.5 Hz

0

600

600

300

200

150

120

100

85.7

75

1

300

300

150

100

75

60

50

42.85

37.5

2

150

150

75

50

37,5

30

25

21.42

18.75

3

75

75

37.5

25

18.75

15

12.5

10.71

9.38

4

37.5

37.5

18.75

12.5

9.38

7.5

6.25

5.36

4.69

5

18.75

18.75

9.38

6.25

4.69

3.75

3.13

2.68

2.34

6

9.38

9.38

4.69

3.13

2.34

1.88

1.56

1.34

1.17

7

4.69

4.69

2.34

1.56

1.17

0.94

0.78

0.67

0.59

Manual DAS 72.1

Page 39