C.4.7 Output Commands – GG, GN, GT, GS, GM, RM, GH, GO, GV, GW
The following commands “Get” the Gross, Net, Tare, long weight and ADC (Sample) values from the DAS
72.1.
GG
Get Gross value
Master ( PC / PLC ) sends
DAS 72.1 responds
Result
GG
¿
G+01.100
Gross 1 100 d
GN
Get Net value
Master ( PC / PLC ) sends
DAS 72.1 responds
Result
GN
¿
N+01.000
Net 1 000 d
GT
Get Tare value
Master ( PC / PLC ) sends
DAS 72.1 responds
Result
GT
¿
T+000.100
Tare 100 d
GS
Get ADC Sample value
This command gets the actual Analogue to Digital Converter (ADC) value. This can be useful during
development or when calibrating to see how much of the ADC range is being used.
Master ( PC / PLC ) sends
DAS 72.1 responds
Result
GS
¿
S+125785
AD-value=125 785 d
For service applications, it is helpful to note the GS values for the “no-load” or “zero” output and when the
“calibration load” is applied.
GM
Get Peak value
This command gets the actual Peak value. This function has to be activated in menu 6.x.1 together with the
command AIx
Master ( PC / SPS ) sends
DAS 72.1 responds
Result
GM
¿
M+01.100
Peak value 1 100 d
Remark
The result of Peak value is depending on setup of Filter FL, Filtermode FM and
Update-Rate UR (average building).
RM
Deletes Peak value
Master ( PC / SPS ) sends
DAS 72.1 responds
Result
RM
¿
OK
Peak value deleted
GH
Get Hold value
Master ( PC / PLC ) sends
DAS 72.1 responds
Result
GH
¿
H+01.100
Hold value 100 d