C.4.7 Output Commands – GG, GN, GT, GS, GM, RM, GH, GO, GV, GW

The following commands “Get” the Gross, Net, Tare, long weight and ADC (Sample) values from the DAS
72.1.

GG

Get Gross value

Master ( PC / PLC ) sends

DAS 72.1 responds

Result

GG

¿

G+01.100

Gross 1 100 d

GN

Get Net value

Master ( PC / PLC ) sends

DAS 72.1 responds

Result

GN

¿

N+01.000

Net 1 000 d

GT

Get Tare value

Master ( PC / PLC ) sends

DAS 72.1 responds

Result

GT

¿

T+000.100

Tare 100 d

GS

Get ADC Sample value

This command gets the actual Analogue to Digital Converter (ADC) value. This can be useful during
development or when calibrating to see how much of the ADC range is being used.

Master ( PC / PLC ) sends

DAS 72.1 responds

Result

GS

¿

S+125785

AD-value=125 785 d

For service applications, it is helpful to note the GS values for the “no-load” or “zero” output and when the
“calibration load” is applied.

GM

Get Peak value

This command gets the actual Peak value. This function has to be activated in menu 6.x.1 together with the
command AIx

Master ( PC / SPS ) sends

DAS 72.1 responds

Result

GM

¿

M+01.100

Peak value 1 100 d

Remark
The result of Peak value is depending on setup of Filter FL, Filtermode FM and
Update-Rate UR (average building).

RM

Deletes Peak value

Master ( PC / SPS ) sends

DAS 72.1 responds

Result

RM

¿

OK

Peak value deleted

GH

Get Hold value

Master ( PC / PLC ) sends

DAS 72.1 responds

Result

GH

¿

H+01.100

Hold value 100 d

Page 42

Manual DAS 72.1