C.4.10 Setpoint Commands - Sn, Hn, Pn, An

The DAS 72.1 has 3 logic outputs where the status is dependent on the weight value (setpoint). Each logic output
can be assigned an independent setpoint value (Sn) with a corresponding hysteresis (Hn), switch action (Pn) and
base (An – switch on the gross or the net weight).

Sn

Setpoint value for logic output 1, 2 or 3

Request / Setting

Master ( PC / PLC ) sends

DAS 72.1 responds

Result

S1

¿

S1:+01500

setpoint S1 = 1 500 d

S1_03000

¿

OK

setpoint S1= 3 000 d

Similarly to read or set the logic output 2 or 3 setpoint, use S2 or S3 instead of S1.

H1

Hysteresis for logic output 1, 2 or 3

Using the H1 command, the hysteresis on the setpoint value is set by the numeric value. Permitted values
are 1 … 9 999.

Master ( PC / PLC ) sends

DAS 72.1 responds

Result

H1

¿

H1:+00100

Hysteresis H1 = 100 d

H1_03000

¿

OK

Hysteresis H1 = 3 000 d

Pn

Switch Logic for output 1, 2 or 3

Using the P1 command, the switch characteristic setpoint S1 ”ON“ or ”OFF“ will be setup. Permitted values
are 0 or 1.

Master ( PC / PLC ) sends

DAS 72.1 responds

Result

P1

¿

P1:+00000

switch logic P1 = 0

P1_00001

¿

OK

switch logic P1 = 1

Example

Setpoint

Hysteresis

Switch logic

Load

Output open

Output closed

S1 = 2 000 kg

H1 = 100 kg

P1 = 1

increasing

³ 2 101 kg

0 ... 2 100 kg

S1 = 2 000 kg

H1 = 100 kg

P1 = 1

decreasing

³ 2 000 kg

1 999 ... 0 kg

S1 = 2 000 kg

H1 = 100 kg

P1 = 0

increasing

0 ... 1 999 kg

³ 2 000 kg

S1 = 2 000 kg

H1 = 100 kg

P1 = 0

decreasing

1 900 ... 0 kg

³ 1 901 kg

Example switch logic P1=1 for setpoint (S1) of 2 000 kg (lines 1 & 2 of table above):

When the weight is increasing between 0 kg and 2100 kg the logic output is “ON”. Once the weight increases
above 2 100 kg then the logic output is “OFF”. The logic output will come “ON” again when the weight value
drops below 2 000 kg.

Example switch logic P1=0 for setpoint (S1) of 2 000 kg (lines 3 & 4 of table above):

When the weight is increasing between 0 kg and 1 999 kg the logic output is “OFF”. Once the weight increases
above 1 999 kg then the logic output is “ON”. The logic output will switch “OFF” again when the weight value
drops below 1 900 kg.

Page 48

Manual DAS 72.1