Factory default settings

This command puts the DAS 72.1 back to a known state. The data will be written to the EEPROM and the
TAC will be incremented by 1.

Note: All calibration and setup information will be lost by issuing this command!

Master ( PC / PLC ) sends

DAS 72.1 responds

Result

CE

¿

E+00017 (example)

Request: TAC-counter CE17

CE_17

¿

OK

Calibration sequence active

FD

¿

OK

Factory default setting

CS

Save the calibration values

This command results in the calibration values being saved to EEPROM, and causes the TAC to be
incremented by 1.

Master ( PC / PLC ) sends

DAS 72.1 responds

Result

CE

¿

E+00017 (example)

Request: TAC-counter CE17

CE_17

¿

OK

Calibration sequence active

CS

¿

OK

Calibration values saved

The CS command saves all of the calibration group values, as set by CZ, CG, CM, DS, DP and ZT. The
command returns ERR and has no updating action unless it is preceded by the CE_XXXXX.