B.3.9 Error codes

Page 16

Manual DAS 72.1

Set final zero out of range.
(You are trying to set a zero which is greater than ±2% of the Upper Display Limit)

Excessive sensitivity requested.
(The input signal is being divided into too many divisions i.e the size of each division is < 0.5 V)

m

Faulty load cell connection (Open circuit / Short Circuit / Wire disconnected)

You are trying to enter a value which is not valid (Outside the acceptable parameter range)

The input signal is in excess of ± 2.6 mV/V

Display underload (Display value is more negative than the lower limit set in Menu 3.1.U)

Display overload (Display value exceeds the upper limit set in Menu 3.1.o)

The zero key is not enabled (See menu 1.1)

Remark
If you need to return the DAS72.1 Mark III to the factory settings simply press and hold the recessed enable switch whilst
powering up the unit. All parameters will then revert to the factory settings

The Zero or Tare motion limit has been exceeded. Set Zero or Tare function disallowed.
Review Zero and Tare motion limits set in menu 4.4