beautypg.com

KEYENCE LK-HD1001 User Manual

Page 55

background image

A-5

Appendices

A

*

If you cannot resume the LK-G5000 Series after taking the above countermeasures or if you
encounter an error which is not listed in the table, contact your nearest KEYENCE office.

Err-62

Parameter range error

The setting value received through the RS-232C
communication is out of the possible setting range.
Check the command you sent.

Err-63

Parameter range error
OUT calculation count
limitation)

The repeated use of an OUT value in OUT
calculation exceeds the limit. Check the command
you sent.

Err-64

Parameter range error
(OUT/Head No.)

The number of sensor heads or OUT being used
exceeds the active head/OUT count.
Check the setting.

Err-65

Parameter range error
(Velocity/acceleration
calculation method)

The OUT set to the measurement type of
"Velocity" or "Acceleration" was set to OUT for
another measurement type or to AVE/P-P/MAX or
other calculation between OUT.
Check the setting.

Err-66

Parameter range error
(OUT specification)

The calculation range set for a certain OUT
includes the OUT itself, or no target OUT has been
set for the AVE/PP/MIN/MAX calculation.

Err-67

Parameter range error
(Sampling cycle)

The specified sampling cycle is faster than the
fastest sampling cycle available based on the
active OUT count, active head count,
measurement mode, and calculation method.
Select a slower sampling cycle or change other
parameters.

Err-68

Parameter range error
(Scaling)

The specified scaling parameters do not satisfy
the following conditions:

• Input value 1 - Input value 2

 0

• |(Displayed value 2 - Displayed value 1) / (Input

value 2 - Input value 1) |

 2.

Err-69

Parameter range error
(Analog output scaling)

The specified analog output scaling parameters do
not satisfy the following conditions:

• Input value 1 - Input value 2

 0

• |(Output voltage value 2 - Output voltage value 1)

/ (Displayed value 2 - Displayed value 1) |

 2.

Err-70

Parameter range error
(Number of data to be
stored)

The specified number of data to be stored exceeds
the possible setting range.

Err-71

Parameter range error
(OUT specified for
data storage)

The number of OUT for which data is stored
exceeds the active OUT count.

Err-88

Timeout error

The delimiting CR expected after receiving a
command via RS-232C communication was not
received for 30 or more seconds.
Check the command you sent and the
communication program.

Err-99

Other error

Contact your nearest KEYENCE office.

Display

System
error

Error description

Countermeasure