English – Compex Performance User Manual
Page 313

311
English
The Compex Performance is designed for use in the electromagnetic
environment stipulated below. The buyer or user of the Compex
Performance must ensure it is used in this recommended environment.
- ELECTROMAGNETIC IMMUNITY
NOTE: VT is the AC supply voltage before application of the test
level.
Electromagnetic
environment - recommendations
Portable and mobile RF communication devices must only be used relative to the Compex
Performance and its wiring at a distance which is not less than the spacing recommended and
calculated using the appropriate equation for the transmitter’s frequency. Recommended
spacing
d = 1.2 √P
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.5 GHz
Where P is the maximum output power of the transmitter in watts (W) set by the manufacturer’s
specifications and where d is the recommended spacing in metres (m). The field intensity of RF fixed
transmitters, as determined by an electromagnetic survey
a
must be less than the observance level
to be found in each frequency range
b
.
Interference may occur close to any appliance identified by the following symbol:
Electromagnetic wave propagation is modified by absorption and reflection due to buildings,
objects and persons.
In the event of abnormal operation, new measures may then be imposed, such as realignment
or movement of the Compex Performance.
b
Above the frequency amplitude from 150 kHz to 80 MHz, the field intensity must be
< 3 V/m
PERFORMANCE FR DE IT ES EN:ENGY Man FR 14/02/08 15:12 Page 311