The Compex Performance is designed for use in the electromagnetic
environment stipulated below. The buyer or user of the Compex
Performance must ensure it is used in this recommended environment.
RECOMMENDATIONS AND DECLARATION BY THE MANUFACTURER -
NOTE: VT is the AC supply voltage before application of the test
level.
Immunity test
Test level IEC 60601
Electrostatic discharge (DES)
IEC 61000-4-2
± 6 kV at the contact
± 8 kV in air
Rapid electrical transients in
bursts of noise
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/output lines
Surge currents
IEC 61000-4-5
± 1 kV differential mode
± 2 kV common mode
Voltage trough, short power
cuts and voltage variations on
power supply input lines
IEC 61000-4-11
< 5% VT (trough > 95% of
VT) for 0.5 cycle
< 40% VT (trough > 60% of
VT) for 5 cycles
< 70% VT (trough > 30% of
VT) for 25 cycles
< 5% VT (trough > 95% of
VT) for 5 seconds
Observance level
± 6 kV at the contact
± 8 kV in air
Not applicable
System battery-powered
Not applicable
System battery-powered
Not applicable
System battery-powered
Magnetic field at mains
frequency (50/60 Hz)
IEC 61000-4-8
3 A/m