The Compex Performance is designed for use in the electromagnetic
environment stipulated below. The buyer or user of the Compex
Performance must ensure it is used in this recommended environment.
RECOMMENDATIONS AND DECLARATION BY THE MANUFACTURER
NOTE: VT is the AC supply voltage before application of the test
level.
Immunity test
Test level IEC 60601
RF immunity
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
10 V/m
26 MHz to 1 GHz
Observance level
Signal lines
Not applicable for power
supplies of devices which are
battery-powered
3 A/m
10 V/m
NOTE 1: From 80 MHz to 800 MHz, the spacing for high frequency amplitude is applied.
NOTE 2: These guidelines may not be appropriate for some situations.
a
The field intensity from fixed transmitters, such as radio telephone base stations
(cellular/wireless) and a mobile radio, amateur radios, AM and FM radio emissions and TV
emissions cannot be predicted with any accuracy. It may therefore be necessary to consider an
analysis of the electromagnetic environment of the site to calculate the electromagnetic
environment coming from RF fixed transmitters. If the field intensity measured in the
environment where the Compex Performance is located exceeds the appropriate RF observance
level above, the Compex Performance should be monitored to ensure it is operating properly.