2 technical data overview, 2 technical data titrator, Technical data overview – BUCHI KjelSampler K-377 User Manual
Page 26: 3 .2 technical data overview

3 Technical data
26
K-375/376/377 Operation Manual, Version B
3 .2
Technical data overview
3 .2 .1
Technical data KjelMaster K-375 and KjelSampler K-376/K-377
Table 3-15: Technical data
KjelMaster K-375
KjelSampler K-376
KjelSampler K-377
Connection voltage
220 – 240 VAC ±10%
100 – 240 VAC ± 10 %
100 – 240 VAC ± 10 %
Frequency
50/60 Hz
50/60 Hz
50/60 Hz
Power consumption
max. 2.2 kW
max. 150 W
max. 150 W
Current consumption (230 V)
9.5 A
650 mA
650 mA
Weight
32 kg
40 kg
(without rack and sample
tubes)
64 kg
(without rack and sample
tubes)
Dimensions (W x H x D)
458 x 670 x 431 mm
505 x 750* x 655
* 1000 mm height
required to allow free
movement of the sampler
arm
1015 x 750** x 655
**1250 mm height
required to allow free
movement of the sampler
arm
Interfaces
RS232
RS232
RS232
Recovery rate
> 99.5 % (1 – 200 mg N)
Reproducibility (RSD)
< 1 %
Measuring range
0.1 – 200 mg N
Environmental conditions
Temperature
Altitude
Humidity
for indoor use only
+ 5 °C to + 40 °C
up to 2000 m above sea level
maximum relative humidity 80 % for temperatures up to 31 °C, decreasing linearly to
50 % relative humidity at 40 °C
Mains connection
Device plug C14
Device plug C14
Device plug C14
Overvoltage category
II
II
II
Pollution degree
2
2
2
Approval
CE/CSA
CE/CSA
CE/CSA
3 .2 .2
Technical data titrator
The following sensors can be connected to the titrator:
• combined pH glass electrode
• optical sensor with measuring range between 50 and 1000 mV (max. 1200 mV)
• temperature measuring sensor for Resistance Thermometer Pt 1000, connection: 2 x 4 mm
sockets and 1 x 2 mm socket
Dosing accuracy:
According to DIN EN ISO 8655, Part 3, or better
Typical accuracy: Systematic error max. ± 30
µL