beautypg.com

2 technical data overview, 2 technical data titrator, Technical data overview – BUCHI KjelSampler K-377 User Manual

Page 26: 3 .2 technical data overview

background image

3 Technical data

26

K-375/376/377 Operation Manual, Version B

3 .2

Technical data overview

3 .2 .1

Technical data KjelMaster K-375 and KjelSampler K-376/K-377

Table 3-15: Technical data

KjelMaster K-375

KjelSampler K-376

KjelSampler K-377

Connection voltage

220 – 240 VAC ±10%

100 – 240 VAC ± 10 %

100 – 240 VAC ± 10 %

Frequency

50/60 Hz

50/60 Hz

50/60 Hz

Power consumption

max. 2.2 kW

max. 150 W

max. 150 W

Current consumption (230 V)

9.5 A

650 mA

650 mA

Weight

32 kg

40 kg

(without rack and sample
tubes)

64 kg

(without rack and sample
tubes)

Dimensions (W x H x D)

458 x 670 x 431 mm

505 x 750* x 655
* 1000 mm height
required to allow free
movement of the sampler
arm

1015 x 750** x 655
**1250 mm height
required to allow free
movement of the sampler
arm

Interfaces

RS232

RS232

RS232

Recovery rate

> 99.5 % (1 – 200 mg N)

Reproducibility (RSD)

< 1 %

Measuring range

0.1 – 200 mg N

Environmental conditions

Temperature
Altitude
Humidity

for indoor use only
+ 5 °C to + 40 °C
up to 2000 m above sea level
maximum relative humidity 80 % for temperatures up to 31 °C, decreasing linearly to
50 % relative humidity at 40 °C

Mains connection

Device plug C14

Device plug C14

Device plug C14

Overvoltage category

II

II

II

Pollution degree

2

2

2

Approval

CE/CSA

CE/CSA

CE/CSA

3 .2 .2

Technical data titrator

The following sensors can be connected to the titrator:

• combined pH glass electrode

• optical sensor with measuring range between 50 and 1000 mV (max. 1200 mV)

• temperature measuring sensor for Resistance Thermometer Pt 1000, connection: 2 x 4 mm

sockets and 1 x 2 mm socket

Dosing accuracy:
According to DIN EN ISO 8655, Part 3, or better
Typical accuracy: Systematic error max. ± 30

µL

This manual is related to the following products: