BUCHI KjelSampler K-377 User Manual
Page 149

8 Troubleshooting
149
K-375/376/377 Operation Manual, Version B
15’114
Sampler: Error during EEPROM writing.
Adjustment value not saved
Call service
17’001
Steam generator overtemperature
Switch distillation unit off and on or call service
17’002
Water level not reached
Check steam generator water supply or call service
18’001
Stirrer out of order
Check stirrer cable or exchange stirrer
50’001
Device powered on
System message, no error.
50’002
Device powered off
System message, no error.
50’003
User login
System message, no error.
50’004
User logout
System message, no error.
50’005
Verify data consistency time
System message, no error.
50‘006
Device power failure during determination
System message, no error.
50’007
Automatic export error
System message, no error.
50‘008
Sequence information
System message, no error.