BUCHI KjelSampler K-377 User Manual
Page 148

8 Troubleshooting
148
K-375/376/377 Operation Manual, Version B
14’100
Titrator timeout
Titration endpoint not reached. Check dosing unit,
if enough titration solution is available or update
frimware.
14’101
Titrator measured value is out of range
Verify the buffers used for calibration, check the
electrode.
14’501
Dosing unit not ready, please check
Verify if the dosing unit is connected to the distillation
unit.
14’502
Dosing unit locked
Check the dosing unit and switch distillation unit off
and on.
14’503
Dosing unit not ready, no exchange unit
Check if dosing unit is connected to the distillation
unit
14’504
Dosing unit not ready, no dosino
Check if dosing unit is connected to the distillation
unit
14’505
Dosing unit overload
Call service
14’506
Dosing unit not ready, cock blocked
Dismantle the dosing unit (see chapter “Trouble-
shooting the dosing unit”)
14’602
Titrator stopped, maximal volume reached
Make sure the electrode is in good working condi-
tions, the correct titration solution is used and that no
air bubbles are in titrator hoses.
14’603
Titrator stopped, stop endpoint reached
Check the titrator and if enough titration solution is
available
14’604
Titrator stopped, stopPot reached
Check the titrator and if enough titration solution is
available
14’605
Titrator stopped, stop time reached
Check the titrator and if enough titration solution is
available
15’001
No sampler connected
Switch on sampler and check the connection cable
15’002
Sampler: Target not achieved
Try again or call service
15’003
Sampler: Connection lost
Check the connection cable
15’101
Sampler: Shield open
Close shield
15’102
Sampler: Crash detected
Try again or call service
15’103
Sampler: Tube not found
Put sample tube in place or call service
15’104
Sampler: Tube not released
Remove sample tube or call service
15’105
Sampler: Error 5, reserve (gibt es nicht)
Call service
15’106
Sampler: Error 6, reserve (gibt es nicht)
Call service
15’107
Sampler: Shield not locked
Call service
15’108
Sampler: Error reference position
Call service
15’109
Sampler zeroadjustment not possible
because x or y deviation is larger then 3mm
or sampler was not in reference position
before adjustment start
Try again or call service
15’110
Sampler: Position error X-axis
Call service
15’111
Sampler: Position error Y-axis
Call service
15’112
Sampler: Position error Z-axis down
Call service
15’113
Sampler: Position error Z-axis up
Call service