8 Troubleshooting
147
K-375/376/377 Operation Manual, Version B
10‘302
Cooling water flow too low
Assure higher flow rate or check parameters in
Settings/Peripherals/Cooling water settings
10‘303
Low pressure during distilling
System pressure is below 150 mbar. Check for leaks
or call service.
10‘311
Pump H2O has no current
Faulty water pump. Exchange pump or call service.
10’312
Pump NaOH has no current
Faulty NaOH pump. Exchange pump or call service.
10’314
Pump H3BO3 has no current
Faulty boric acid pump. Exchange pump or call
service.
12’001
Valve steam (Y1) out of order
Faulty valve or wire harness. Call service.
12’002
Valve cooling water in (Y5) out of order
Faulty valve or wire harness. Call service.
12’003
Valve sampler steam (Y6) out of order
Faulty valve or wire harness. Call service.
12’004
Valve sampler transfer (Y7) out of order
Faulty valve or wire harness. Call service.
12’005
Valve 5 (not used) out of order
Faulty valve or wire harness. Call service.
12’006
Valve waste sample tube (Y2) out of order
Faulty valve or wire harness. Call service.
12’007
Valve aspiration in (Y3) out of order
Faulty valve or wire harness. Call service.
12’008
Valve receiver (Y4) out of order
Faulty valve or wire harness. Call service.
12’009
Valve H2O injection (Y8) out of order
Faulty valve or wire harness. Call service.
12’010
Valve H2O sample tube (Y9) out of order
Faulty valve or wire harness. Call service.
12’011
Valve waste receiver (Y10) out of order
Faulty valve or wire harness. Call service.
13’001
27V power supply overcurrent
Faulty electronic board. Call service.
13’002
Ventilator power overcurrent
Fan short circuit. Call service.
13’003
Ventilator electronics blocked
Check for blockage or call service
13’004
Ventilator interior blocked
Check for blockage or call service
14’001
Titrator not ready
Check if all cables are plugged to the titrator, restart
the system or call service.
14’002
Titrator information (version)
Call service
14‘003
Titrator not started
Titrator error. Check function in System Preparation/
Buret function. Call service.
14‘004
Titrator not started, pH value too low
pH value is below the set endpoint. Verify the elec-
trode, dosing unit and boric acid.
14‘005
Titrator not started, pH value too high
pH value is above the set endpoint. Verify the elec-
trode, dosing unit and receiving solution.
14’006
Wrong titration direction
Ensure pH electrode is imersed in receiving solution
and verify if the right titration solution is used.
14’007
Titration speed over specification
Overtitration. Use lower concentrated titration solution
or reduce titration speed.
14’008
Titration speed over specification
Overtitration during back titration. Use lower concen-
trated titration solution or reduce titration speed.
14’010
Titrator module could not create service 11
Distillation unit needs to be switched off and on
14’011
Titrator module could not create service 21
Distillation unit needs to be switched off and on
14’012
Titrator module could not create service 41
Distillation unit needs to be switched off and on
14’013
Titrator module could not create service 3
Distillation unit needs to be switched off and on