BUCHI KjelSampler K-377 User Manual
Page 110

6 Operation
110
K-375/376/377 Operation Manual, Version B
Result units
Create a new unit with
NEW and store it under a
new, unique name.
Select a base unit for the new unit calculation
Selectable are:
g N/L, % N, g Pr/L or % Pr
(N = nitrogen content, Pr = protein content)
Enter a correction factor and/or correction offset
if necessary and select the preferred number of
decimal places.
(The results are calculated following the linear equa-
tion
y= a·x+b where a is given by the correction
factor and
b by the correction offset.)
Dosage volume in status view
Set the increment in
[mL/click] for the direct
dosing of H
2
O, NaOH and H
3
BO
3
within the status
view. (Per click on the corresponding dose button,
the specified amount in mL will be dosed.)