Rice Lake MC Series Mass Comparator, A&D Weighing - MC-30K User Manual
Page 55

53
Display
Error code Description
Automatic response adjustment zero error 
The automatic response adjustment can not be performed 
because there is something on the pan. Clear the pan. Press 
the 
CAL
key to return to the weighing mode.
(Check NG)
Automatic response adjustment unstable error 
The automatic response adjustment can not be performed 
because the weight value is unstable. 
Check the ambient conditions such as breeze, vibration
and magnetic fields, also check the weighing pan.
Press the
CAL
key to return to the weighing mode.
(Check no)
Internal error 
This error indicates an internal error as the result of the 
self-check function. Repair is required. Contact the local A&D 
dealer. 
Clock battery error 
The clock backup battery has been depleted. 
Press any key and set the time and date. The clock and 
calendar function works normally as long as the AC adapter 
is connected to the balance. If this error appears frequently, 
contact the local A&D dealer. 
(Blinking)
Memory full 
The amount of weighing data in memory has reached the 
maximum capacity. Delete the data in memory to store new 
data. For details, refer to the GX-K manual "12. Data Memory". 
(Illuminated)
Memory full 
The amount of calibration or calibration test data in memory 
has reached the maximum capacity (50 sets). 
The data in memory will be deleted automatically to store 
new data. For details, refer to the GX-K manual "12. Data 
Memory". 
Memory type error 
The type of memory set in the function table and the type of 
data stored are different. For details, refer to the GX-K manual 
"12. Data Memory". 
EC,E00
Communications error 
A protocol error occurred in communications. 
Confirm the format, baud rate and parity. 
EC,E01
Undefined command error 
An undefined command was received. 
Confirm the command. 
EC,E02
Not ready 
A received command can not be processed. 
Example: 
The balance received a "Q" command, but not in the
weighing mode.
The balance received a "Q" command while processing a
RE-ZERO command.