15-2. Error Codes
Display
Error code Description
EC,E11
Stability error 
The balance can not stabilize due to an environmental 
problem. Check around the pan. Refer to "3. Precautions". 
Prevent vibration, drafts, temperature changes, static 
electricity and magnetic fields, from influencing the balance. 
To return to the weighing mode, press the 
CAL
key.
Out of the setting range 
The data to be stored is out of the setting range. 
EC,E16
Internal mass error 
Applying the internal mass does not yield a change in the 
weight value as specified. 
Confirm that there is nothing on the pan and perform the 
weighing operation from the beginning again. 
EC,E17
Internal mass error 
The internal mass application mechanism does not function 
properly. Perform the weighing operation from the beginning 
again. 
EC,E20
Calibration weight error 
The calibration weight is too heavy. Confirm the calibration 
weight value. Press the 
CAL
key to return to the weighing
mode.
EC,E21
Calibration weight error 
The calibration weight is too light. Confirm the calibration 
weight value. Press the 
CAL
key to return to the weighing
mode.
Overload error 
A sample beyond the balance weighing capacity has been 
placed on the pan. Remove the sample from the pan. 
Weighing pan Error 
The weight value is too light. 
Confirm that the weighing pan is properly installed and 
calibrate the balance. 
Sample mass error 
The balance can not store the sample for the counting mode 
or for the percent mode because it is too light. 
Use a larger sample. 
Unit mass error 
The sample unit mass for the counting mode is too light. 
Storing and using it for counting will cause a counting error. 
Add samples to reach the specified number and press the 
key. Pressing the
key without adding samples
will shift the balance to the counting mode. But, to acquire 
accurate weighing, be sure to add samples. 