9-3. Description of the Class "Environment, Display"

Condition ( Cond )

Cond 0

This parameter is for sensitive response to the fluctuation of a weight value. Used
for powder target weighing, weighing a very light sample or when quick response
weighing is required. After setting, the balance displays

FAST

.

Cond 2 This parameter is for stable weighing with slow response. Used to prevent a

weight value from drifting due to vibration or drafts. After setting, the balance
displays

SLOW

.

Notes In automatic response adjustment, the weighing speed is selected

automatically.

With "Hold function (

Hold

)" set to "ON (

1

)", this item is used to set the

averaging time.

Stability band width ( 5t-b )

This item controls the width to regard a weight value as a stable value. When the fluctuation
per second is less than the parameter, the balance displays the stabilization indicator and
outputs or stores the data. The parameter influences the "Auto print mode"

5t-b

0 This parameter is used for sensitive response of the stabilization indicator. Used

for exact weighing.

5t-b

2 This parameter ignores slight fluctuation of a weight value. Used to prevent a

weight value from drifting due to vibration or drafts.

Note

With "Hold function (

Hold

)" set to "ON (

1

)", this item is used to set the

stabilization range.

Zero upon power-on ( p-tr )

When a hopper is attached to the weighing pan and loss-in weighing is performed, the
remaining amount of the material will become unknown if tare is performed each time a
weighing starts.
When " p-tr " is set to “ 1 ”, tare is not performed at weighing start. So, the remaining amount of
the material can be monitored, when the power is turned on again after it was turned off.

Span calibration ( 5pn )

When a hopper is attached to the weighing pan and calibration is to be performed with the
hopper attached, set " 5pn " to “ 1 ”. When the tare value (hopper and other devices attached) is
within the value in the table below, calibration using the internal mass is possible.

Model

Tare

value

MC-10K

9 kg or less

MC-30K

29 kg or less