Rice Lake MC Series Mass Comparator, A&D Weighing - MC-30K User Manual
Page 31

29
Class
Item and Parameter
Description
Unit
Unit
Refer to GX-K manual "5. Weighing Units".
C5 in
Internal mass correction
Displayed only when the internal mass value correction switch is set
to 1. Refer to "7. Calibration".
id
ID number setting
Refer to "10. ID Number And GLP Report".
0
1
2
3
4
5
6
f1-b
Averaging range for
the first moving
average
7
Small
Large
0
No averaging
1
0.5 second
2
1.0 second
3
1.5 seconds
4
2.0 seconds
5
2.5 seconds
6
3.2 seconds
7
4.8 seconds
f1-t
Averaging time for
the first moving
average
8
6.4 seconds
0
1
2
3
4
5
f2-b
Averaging range for
the second moving
average
6
Small
Large
0
No averaging
1
0.5 second
2
1.0 second
3
1.5 seconds
4
2.0 seconds
5
2.5 seconds
erfnc
Extended
function
f2-t
Averaging time for
the second moving
average
6
3.2 seconds
: Factory settings. Digit is a unit of minimum weighing value.
Caution
The balance may not transmit the data completely at the specified refresh rate, depending
on the baud rate or data added to the weighing data such as time, date and ID number.