7-5. Calibration Test Using an External Weight

This function tests the weighing accuracy using an external
weight and outputs the result. This is available only when
the "GLP output (

info

)" parameter is set to "

1

" or "

2

".

(Calibration test does not perform calibration)

When the MC-10K is used

Step 1 Connect the AC adapter and warm up the balance

for at least 30 minutes with nothing on the weighing
pan.

Step 2 Press and hold the

CAL

key until

CCout

is

displayed, then release the key.

Step 3 The balance displays

CC 0 .

If you want to change the target weight, press

the

SAMPLE

key and go to step 4.

A list of usable weights is shown on page 17.

If you use the target weight value stored in

the balance, go to step 5.

Step 4 Specify the target weight value as follows:

SAMPLE

key... To To switch between the calibration weight

selection mode (All of the digits blinking) or
the value adjustment mode (The selected
digits blinking).
With the MC-10K, the digits in the 1

st

, 2

nd

,

and 3

rd

decimal places can be adjusted.

With the MC-31K, the digits in the 1

st

and 2

nd

decimal places can be adjusted.

RE-ZERO

key . To select the target weight or adjust the

*value. In the value adjustment mode, -15
digits appear after +15 digits.

PRINT

key...... To store the new weight value. Even if the

AC adapter is removed, the data is
maintained in non-volatile memory.

CAL

key ......... To cancel the operation and return to

CC 0

.

Step 5 Confirm that there is nothing on the pan and press the

PRINT

key. The balance measures the zero point and displays the
measured value. Do not allow vibration or drafts to affect the
balance. The balance displays the target weight value.

Step 6 Place the displayed target weight on the pan and press the

PRINT

key. The balance measures the target weight and

displays the measured value. Do not allow vibration or
drafts to affect the balance.

Step 7 The balance displays

end

.

Remove the weight from the pan.

Step 8 The balance displays

glp

and outputs the

"calibration test report" using the RS-232C interface
or stores the calibration test data in memory. Refer to
"10-2. GLP Report" of the function table for details.

Step 9 The balance will automatically return to the weighing mode.

Release

Press and hold

Select

Place weight

Remove

GLP output

Example:

New weight

2000.123 g

Press

C C 0

C al i n

0.0 00

g

C C o ut

0.0 00

g

en d

gl p

en d

2

k

2

k

2 00 0.12 3

g

C C 0

0 .00 0

g

0 00 0.0 00

g

1

2 00 0.0 00

g

2 00 0.1 20

g

2 00 0.0 00

g

2 00 0.1 20

g

20 00 .1 23

g