6-2. Self Check Function With Response Adjustment

This function automatically updates the response adjustment by analyzing the influence of
the environment on the weighing data and also self-checks the balance performance using
the internal mass.


Step 1 Press and hold the

MODE

key until

RESPONSE

is

displayed, and then release the key.

Step 2 The balance automatically starts to check the balance

performance and sets the response characteristic.

Caution Do not allow vibration or drafts to affect the

balance during adjustment.

Step 3 After automatic adjustment, the balance displays the updated

response indicator and returns to the weighing mode.

The response indicator remains displayed for a while.

Example of display

MID.

and

OK

: The example above indicates that the result of the self check

is good and MID. is selected as the response rate..

Note

If improper performance is found in the self check, the balance displays

CH no

.

Contact the local A&D dealer for repair.

If the automatic response adjustment fails, the balance displays

CH ng

. Check the

ambient conditions such as breeze and vibration, also check the weighing pan. Then,
perform the adjustment again. To return to the weighing mode, press the

CAL

key.


C H . . . .

C H

0.0 00

g

Result

Press and hold

Release

This manual is related to the following products: