beautypg.com

2. glp report, Calibration report using the internal mass – Rice Lake MC Series Mass Comparator, A&D Weighing - MC-30K User Manual

Page 39

background image

37

10-2. GLP Report

Set the following parameters to output the report.

To print the report, set the "GLP output (

info

)" parameter to "

1

" and use MODE 3 of the

AD-8121B. Refer to "11-2-1. Connection to the AD-8121B Printer" for details on using the
printer.

To output the report to a personal computer using the RS-232C interface, set the "GLP

output (

info

)" parameter to "

2

".

If the time and date are not correct, set the correct time and date in "Clock (

Cl adj

)" of the

function table.


Notes

For operational details about calibration and calibration test, refer to "7. Calibration".


Calibration report using the internal mass

Setting of "info 1"

Setting of "info 2"

AD-8121 printer format

General format













Space, ASCII 20h

Terminator, CR , LF or CR

CR

Carriage return, ASCII 0Dh

LF

Line feed, ASCII 0Ah

Manufacturer

Model

Serial number

ID number

Date

Time

Calibration type

Signature

===========

A

=

&

=

D

MODEL

=====

MC-30K

S/N

=====

01234567

ID

=======

ABCDEFG

DATE

======

2009/12/31

TIME

========

12:34:56

CALIBRATED(INT.)

SIGNATURE

- - - - - - - -

~~~~~~~~~~~A~&~D
MODEL~~~~~MC-30K

S/N~~~~~01234567

ID~~~~~~~ABCDEFG

DATE

~~~~~~2009/12/31
TIME

~~~~~~~~12:34:56
CALIBRATED(INT.)

SIGNATURE



----------------


This manual is related to the following products: