beautypg.com

2. details of the function table – Rice Lake MC Series Mass Comparator, A&D Weighing - MC-30K User Manual

Page 28

background image

26

9-2. Details of the Function Table

Class

Item and Parameter

Description

0

Fast response, sensitive value

FAST

1

MID.

Cond

Condition

2

Slow response, stable value

SLOW

Can be changed by response

adjustment. With "Hold 1", sets the

averaging time.

0

Stable when within ±1 digit

1

5t-b

Stability band width

2

Stable when within ±3 digits

The stabilization indicator illuminates

with the display fluctuation within the

range. With "Hold 1", sets the stable

range.

0

OFF

Hold

Hold function

1

ON

Holds the display when stable in

animal mode. With "

Hold 1

",

ANIMAL

turns on.

0

OFF

1

Normal

2

Strong

trc

Zero tracking

3

Very strong

Keeps zero display by tracking zero
drift.

0

5 times/second

5pd

Display refresh rate 1

10 times/second

Period to refresh the display

0

Point (.)

pnt

Decimal point

1

Comma (,)

Decimal point format

0

OFF

p-on

Auto display-ON

1

ON

Turns on the weighing mode display
when the AC adapter is connected.

0

OFF

poff

Auto display-OFF

1

ON (10 minutes)

Turns off the display after 10
minutes of inactivity.

0

OFF

g5i

Capacity indicator

1

ON

Capacity indicator.

Zero:

0%

Maximum capacity: 100%

0

OFF

add

Accumulation function 1

ON

Displays and outputs the total value
of the weighing data.

0

Does not display

rng

Display at start

1

Displays

Select whether or not to display the

smallest displayable weighing value

at weighing start.

0

Not used

f1l

Filter

1

Used (when the balance is used as a mass comparator)

0

Sets the display to zero.

p-tr

Zero upon power-on

1

Does not set the display to zero. Displays the previous value.

With this setting, do not perform re-zero operations frequently.

Refer to“Zero upon power-on” on page 30.

0

Performs zero and span calibration

ba5fnc
Environment
Display

5pn

Span calibration

1

Performs span calibration only.

Span calibration using the internal mass is possible with a tare on

the weighing pan. Refer to “Span calibration” on page 30.

Cl adj
Clock

Refer to the GX-K manual “10-7. Clock and Calendar Function"

Confirms and sets the time and

date. The time and date are added

to the output data.

0

No comparison

1

Comparison, excluding "near zero" when stable value or overloaded

2

Comparison, including "near zero" when stable value or overloaded

3

Continuous comparison, excluding "near zero"

Cp

Comparator mode

4

Continuous comparison, including "near zero"

0

Set the upper lower limit value

Select Cp Hi or Cp lo.

Cp in

Data input method

1

Set the reference value

Select Cp ref or Cp lmt.

0

Not added

Cp fnc
Comparator

Cp-r

Comparison results 1

Added

Select whether or not to add the

comparison results to the output

data.

: Factory settings. Digit is a unit of minimum weighing value.

This manual is related to the following products: