2. automatic self calibration – Rice Lake MC Series Mass Comparator, A&D Weighing - MC-30K User Manual

Page 19

background image

17

Caution on using an external weight

The accuracy of an external weight can influence the accuracy of weighing. Select an

appropriate weight as listed below:

Select a mass for calibration and calibration test from the following table.

Model Usable

calibration

weight Adjustable

range

MC-10K

2kg, 3kg, 4kg, 5kg, 6kg, 7kg, 8kg, 9kg, 10kg

-0.150g ~ +0.159g

MC-30K

20kg, 30kg

-1.50g ~ +1.59g

The calibration weight in bold type: factory setting

The calibration weight value can be adjusted within the range above.

Display

This indicator means "In process of measuring calibration data".
Do not allow vibration or drafts to affect the balance while the
indicator is displayed.


7-2. Automatic Self Calibration

Automatic self calibration due to changes in temperature

This function automatically calibrates the balance when the balance detects an ambient

temperature change. If GLP output is selected in the function table, the balance outputs the
calibration report or stores the data in memory. Automatic self calibration functions even if
the display is turned off (standby state). Refer to "8-1. Permit Or Inhibit" for the operation.

Caution

If something is on the weighing pan, the balance judges that it is in use and does not

perform automatic self calibration. To maintain the calibrated state, keep the
weighing pan clear while not in use.

Indicates that the balance detects a change in ambient temperature and
automatic self calibration will start. If the balance is not used for a few
minutes with this indicator blinking, the balance performs automatic self
calibration. The blinking duration depends on the environment.

Indicates that the balance is measuring calibration data. Do not allow
vibration or drafts to affect the balance while this indicator is displayed.
After calibration, the balance returns to indicate the previous display.

Note The balance can be used while the indicator blinks. But, it is recommended that to

maintain the accuracy, stop using the balance and confirm that there is nothing on the
pan and allow the balance to perform self calibration.