8-2. Initializing the Balance

This function returns the following parameters to factory settings.

Calibration

data

Function

table

The sample unit mass value (counting mode),

100% reference mass value (percent mode)

The data that is stored in the balance using the data memory function
External calibration weight and target weight value
Function switch settings


Note Be sure to calibrate the balance after initialization.

Step 1 Press the

ON:OFF

key to turn off the display.

Step 2 While pressing and holding the

PRINT

key and

the

SAMPLE

key, press the

ON:OFF

key to

display

p5

.

Step 3 Press the

SAMPLE

key to display

Clr

.

Step 4 Press the

PRINT

key.

To cancel this operation, press the

CAL

key.

Step 5 Press the

RE-ZERO

key to display

Clr go

.

Step 6 Press the

PRINT

key to initialize the balance.

The balance will automatically return to the weighing
mode.

en d

0 .0 0 0

g

Cl r n o

Cl r

p5

Cl r g o

With these keys Press the key

held down