47

13. Extended Function

The MC series balance has several extended functions equipped for special applications or to
troubleshoot when using the standard functions.

When the "Filter ( f1l)” is set to “ 0 “, extra items are available as shown below.

Class

Item and Parameter

Description

0

1
2
3

4

5
6

f1-b

Averaging range for

the first moving

average

7

Small

Large

When the fluctuation of a weight value is within

the averaging range, the averaging operation

starts to stabilize the displayed value.

When the fluctuation of a weight value is

small, for example, when weighing or filling a

small amount of sample, the averaging is

always performed and the response rate

maybe slow. Under such a condition, change

the parameter.

Refer to “Averaging range

( f1-b )

and

averaging time

( f1-t )”

0

No averaging

1

0.5 second

2

1.0 second

3

1.5 seconds

4

2.0 seconds

5

2.5 seconds

6

3.2 seconds

7

4.8 seconds

f1-t

Averaging time for

the first moving

average

8

6.4 seconds

When the fluctuation of a weight value is within
the averaging range, the averaging operation
starts. Once the duration of averaging operations
reach the averaging time, moving averaging
starts. This parameter sets the time to start
moving averaging.
Refer to “Averaging range ( f1-b ) and
averaging time ( f1-t )”

0

1

2
3
4
5

f2-b

Averaging range for

the second moving

average

6

Small

Large

Refer to “Filter depending on differences in
the amount to deliver powdery and liquid
material”.

0

No averaging

1

0.5 second

2

1.0 second

3

1.5 seconds

4

2.0 seconds

5

2.5 seconds

erfnc

Extended

function

f2-t

Averaging time for

the second moving

average

6

3.2 seconds