Rice Lake MC Series Mass Comparator, A&D Weighing - MC-30K User Manual
Page 51

49
13-1-2. Filter depending on differences in the amount to
deliver powdery and liquid material
z
When the weight value is not stable even without load fluctuations
Increase the averaging range for the first moving average parameter ( f1-b ).
Increase the averaging time for the first moving average parameter ( f1-t ).
Strengthen the digital filter. (Increase the function setting “Condition ( Cond )” parameter.)
z
When response is slow during full and medium flow
Decrease the averaging range for the first moving average parameter ( f1-b ).
Weaken the digital filter. (Decrease the function setting “Condition ( Cond )” parameter.)
z
When response is slow during dribble flow
Decrease the averaging time for the first moving average parameter (f1-t).
Weaken the digital filter. (Decrease the function setting “Condition ( Cond )” parameter.
z
To increase stability without load fluctuations
Increase the averaging range for the second moving average parameter ( f2-b ).
Increase the averaging time for the second moving average parameter ( f2-t ).