28

Class

Item and Parameter

Description

0

No output

d-no
Data number output

1

Output

Refer to the GX-K manual "12.

Data Memory"

0

No output

1

Time only

2

Date only

5-td
Time/Date output

3

Time and date

Selects whether or not the time

or date is added to the weighing

data. Refer to the GX-K manual

"10-7. Clock and Calendar

Function" for details.

0

No output

5-id
ID number output

1

Output

Selects whether or not the ID

number is output.

0

No pause

pU5e
Data output pause

1

Pause (1.6 seconds)

Selects the data output interval.

0

Not used

at-f
Auto feed

1

Used

Selects whether or not automatic

feed is performed.

0

No output

1

AD-8121 format

info
GLP output

2

General data format

Selects GLP output method.

For how to set time and date to

be added, refer to the GX-K

manual "10-7. Clock and

Calendar Function".

0

Not used

dout
Data output

ar-d
Zero after output

1

Used

Adjusts zero automatically after

data is output

0

600

bps

1

1200

bps

2

2400

bps

3

4800

bps

4

9600

bps

bp5
Baud rate

5

19200 bps

0

7 bits, even

1

7 bits, odd

btpr
Data bit, parity bit

2

8 bits, none

0

CR LF

Crlf
Terminator

1

CR

CR: ASCII code 0Dh

LF: ASCII code 0Ah

0

A&D standard format

1

DP format

2

KF format

3

MT format

4

NU format

type
Data format

5

CSV format

Refer to "9-5. Description of Item

"Data Format".

0

No limit

t-Up
Timeout

1

1 second

Selects the wait time to receive a

command.

0

No output

erCd
AK, Error code

1

Output

AK: ASCII code 06h

0

Not used

5if
Serial
interface

Ct5
CTS, RTS control

1

Used

Controls CTS and RTS.

mlt
Programmable-unit (Multi-unit)

Available only when programmable-unit mode is selected.

Refer to the GX-K manual "14. Programmable Units" for details".

: Factory settings. Digit is a unit of minimum weighing value.

This manual is related to the following products: