beautypg.com

3. control using cts and rts, 4. settings related to rs-232c – Rice Lake MC Series Mass Comparator, A&D Weighing - MC-30K User Manual

Page 48

background image

46


R command

















When a communication error has occurred due to external noise, or a parity error has

occurred due to transmission error, the balance transmits an error code. In this case, send
the command again.


xx is error code number.

12-3. Control Using CTS and RTS

Depending on the "Ct5" parameter of "Serial interface (5if)", the balance performs as follows:

Ct5 0

Regardless of whether the balance can receive a command or not, the balance keeps the
CTS line HI. The balance outputs data regardless of the condition of the RTS line.

Ct5 1

The CTS line is kept Hi normally. When the balance can not receive the next command
(Example: while the balance is processing the last command), the balance sets the CTS line
to Lo. The balance confirms the level of the RTS line before outputting a set of data. If the
RTS level is Hi, the balance outputs data. If the RTS level is Lo, data is not output (The data
is canceled).


12-4. Settings Related to RS-232C

Concerning the RS-232C, the balance has two functions: "Data output (

dout

)" and "Serial

interface (

5if

)". Set each function as necessary.


Command received

Command completed

Weighing mode

Zero display

In process

Personal computer Time Balance

R command

R

CR LF

AKCR LF

AKCR LF