Fig. 8: hampshire tsharc control panel – Kontron Medi Client IIA 104 (EOL) User Manual
Page 23

7. Product Description
Medi Client IIA - Instructions for use (Version 1.03)
7.1.3.1.
Calibrating the Touch Screen
Calibration serves two purposes:
Sets the active area of the touch screen
Aligns the active area of the touch screen to the screen’s image.
Before you calibrate the touch screen, let the system warm up for 30 minutes.
Calibration aligns the active touch-sensitive area of the touch screen with the image on the display. Calibration also
determines the edges of the screen’s image and locates the center of the touch screen. If the touch screen is not
calibrated properly, the active area of the touch screen may not be aligned with the screen’s image or may be
unnecessarily small in size. If necessary, in order to access the calibration routine, use an USB mouse.
7.1.3.2.
Calibrate the Touch Screen
The touch screen of your Medi Client IIAsystem is factory calibrated. Run the calibration routine when an alignment
problem exists between the mouse pointer and the stylus contact location on the screen. You can adjust the touch screen
calibration by:
Running the touch screen property sheet from the Start-Programs and clicking on “Hampshire TSHARC Control Panel”.
The following screen will appear:
Fig. 8: Hampshire TSHARC Control Panel
Navigate using the available tabs.
www.kontron.com
21