Kontron Medi Client IIA 104 (EOL) User Manual
Page 15

5. General Safety Instructions
Medi Client IIA - Instructions for use (Version 1.03)
Recommended separation distances between portable and mobile RF communications equipment and the
Medi Client IIA
The Medi Client IIA is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the Medi Client IIA can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the Medi Client IIA as recommended below, according
to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter
[m]
Rated maximum output power
of transmitter
[W]
150 kHz to 80 MHz
P
d
4
,
0
=
80 MHz to 800 MHz
P
d
4
,
0
=
800 MHz to 2,5 GHz
P
d
7
,
0
=
0,01
0,04 0,04 0,07
0,1
0,13 0,13 0,22
1
0,40 0,40 0,70
10
1,26 1,26 2,21
100
4,00 4,00 7,00
For transmitters rated at a maximum output power not listed above, the recommended separation
distance
d in metres [m] can be estimated using the equation applicable to the frequency of the transmitter,
where
P is the maximum output power rating of the transmitter in watts [W] according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
Table 4: Recommended separation distances between portable and mobile RF communications equipment and the Medi Client IIA
www.kontron.com
13