Compliant accessories, 1 “compliant accessories – Kontron Medi Client IIA 104 (EOL) User Manual
Page 16

5. General Safety Instructions
Medi Client IIA - Instructions for use (Version 1.03)
Attention:
Use of mobile phones or other HF power transmitting devices nearby the Medi Client IIA maybe influence the function and
correct visualization of signals. Therefore please switch off any HF energy emitting devices or keep a distance of minimum
3 m (10 feet) to the Medi Client IIA system. Please observe the value included in the “Table 4”.
Caution:
This device works without deviation and the safe operation is ensured only, if the safety, the EMC and the ESD precautions
contained in the “Instructions for Use” are observed by the user. Also the labels, stickers and markings on device itself
shall be taken into consideration. Please observe the warnings and follow the instructions contained in the “Instructions
for Use” or on the labels of the device.
Don't stack this device with other electronic devices. Interacting EMC influence may be happen.
CAUTION
The device is not Defibrillation proved! Be carful, system maybe destroyed by disregarding of this hint.
CAUTION
Possible fire hazard, explosion or ignitions will be happen, when used this device in oxygen enriched
atmosphere or mixtures of oxygen and flammable vapors, such as alcohol, ether, ethylene and
cyclopropane.
The device is not AP or APG classified!
Safe Networking with Medi Client IIA
For a safe networking with electrical equipment, the user is responsible to ensure that the resulting system meets the
requirements set forth by the following standards for medical equipment:
EN 60601-1 (IEC 60601-1)
Part 1: General requirements for safety
EN 60601-1-1 (IEC 60601-1-1) Part 1-1: General requirements for safety;
Safety requirements for medical electrical systems
EN 60601-1-2 (IEC 60601-1-2) Part 1-2: General requirements for safety
Electromagnetic compatibility; Requirements and tests
EN 60601-1-4 (IEC 60601-1-4) Part 1-4: General requirements for safety
Programmable electrical medical systems
5.4.1. Compliant Accessories
Shielded serial cable, length up to 3 m (118.11”)
10/100/1000 CAT6 Ethernet cable, length up to 10 m (393.7”)
USB Mouse, Microsoft Wheel Mouse Optical 1.1A USB
USB Keyboard Cherry, Model RS6500 USB
AC/DC- Adapter, Kontron Part Number (PN): 1036-2985
For VESA
®
75/100 mounting: M4x8 zinc-coated Phillips screw (PN: 0-0068-1058) for or M4x8 V2A stainless steel, Allen
screw, flat-head (PN: 0-0068-2575)
14